User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: Validyne MC1-3:Three Channel Module Case INSTRUMENT: Validyne MC1-20:Twenty Channel Module Case DATE: 24-Nov-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 24 NUMBER OF LINES: 197 CONFIGURATION: Datron 1281 STANDARD: Clarostat 240-C ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 STD Clarostat 240-C 1.004 IEEE [@1281]*RST 1.005 HEAD INITIAL CONDITIONS 1.006 DISP The test adaptor consists of two cards. 1.006 DISP One labled VOLTMETER and the othe labled LOAD. 1.007 DISP When directed to make connections for the DVM, these 1.007 DISP connections MUST be made on the card labled VOLTMETER. 1.008 DISP When directed to make connections for the Clarostat, 1.008 DISP these connections must be made on the card labled LOAD. 1.009 DISP Plug the test adaptor in to one of the UUT slots. 1.010 DISP Plug the UUT power cord into to the variac output. 1.011 DISP Turn on the variac and set its output for 115 volts . 1.012 DISP Turn on the UUT. 1.013 HEAD {* DC Power Supply Tests *} 1.014 HEAD { Load Regulation } 1.015 JMP 2.003 1.016 EVAL 2.001 DISP Remove the Clarostat from the test adaptor. 2.002 JMP 2.004 2.003 DISP Connect the Datron 1281 to the GRD. and +15 terminals. 2.004 IEEE [@1281]DCV AUTO 2.005 IEEE [@1281]RDG?[I] 2.006 MEME 2.007 JMPT 2.010 Validyne MC1-20:Twenty Channel Module Cas 2.008 DISP Set the Clarostat to 100 ohm and connect it to the GRD 2.008 DISP and +15 terminal posts. 2.009 JMP 2.011 2.010 DISP Set the Clarostat to 30 ohm and connect it to the GRD 2.010 DISP and +15 terminal posts. 2.011 IEEE [@1281]RDG?[I] 2.012 MEME 2.013 MEMC V 0.100% 3.001 JMP 4.003 3.002 EVAL 4.001 DISP Remove the Clarostat from the -15 terminal. 4.002 JMP 4.004 4.003 DISP Move the Datron HI connection to the -15 terminal post. 4.004 IEEE [@1281]RDG?[I] 4.005 MEME 4.006 DISP Move the Clarostat hi connection to the -15 terminal. 4.007 IEEE [@1281]RDG?[I] 4.008 MEME 4.009 MEMC V 0.100% 5.001 HEAD { Line Regulation } 5.002 JMP 6.001 5.003 EVAL # 2.002 DISP Connect the Datron 1281 to the +15 and GRD. terminals # 2.002 DISP on the test adaptor. 6.001 IEEE [@1281]DCV AUTO 6.002 IEEE [@1281]RDG?[I] 6.003 MEME >1 6.004 DISP Set the variac for 104 volts. 6.005 IEEE [@1281]RDG?[I] 6.006 MEME 6.007 MEME <1 6.008 MEMC V 0.100% 104VAC #! T.U.R. not calculated because System Uncertainty not available. 7.001 DISP Set the variac for 127 volts. 7.002 IEEE [@1281]RDG?[I] 7.003 MEME 7.004 MEME <1 7.005 MEMC V 0.100% 127VAC #! T.U.R. not calculated because System Uncertainty not available. 8.001 DISP Set the variac for 115 volts. 8.002 DISP Move the Datron HI connection to the +15 terminal post. 8.003 DISP Move the Clarostat hi connection to the +15 terminal. 8.004 JMP 9.002 8.005 EVAL 9.001 DISP Reset the variac for 115 volts 9.002 IEEE [@1281]DCV AUTO 9.003 IEEE [@1281]RDG?[I] 9.004 MEME >1 9.005 DISP Set the variac for 104 volts. 9.006 IEEE [@1281]RDG?[I] 9.007 MEME 9.008 MEME <1 9.009 MEMC V 0.100% 104VAC #! T.U.R. not calculated because System Uncertainty not available. 10.001 DISP Set the variac for 127 volts. 10.002 IEEE [@1281]RDG?[I] 10.003 MEME 10.004 MEME <1 10.005 MEMC V 0.100% 127VAC #! T.U.R. not calculated because System Uncertainty not available. 11.001 DISP Set the variac for 115 volts. 11.002 DISP Disconnect the Clarostat from the UUT. 11.003 HEAD {* Carrier Tests *} 11.004 JMP 12.001 11.005 EVAL 12.001 DISP Connect the Datron to the AC terminals on the adaptor. 12.001 DISP Make all connections to the card inside the UUT. 12.002 IEEE [@1281]ACV AUTO 12.003 ACC 5.000V 0.01% 12.004 IEEE [@1281]RDG?[D2000][I] 12.005 MEME 12.006 MEMC V 0.100U #! Test Tol 0.1, Sys Tol 0.0005, TUR 200.000 (>= 4.00). 13.001 IEEE [@1281]FREQ?[D2000][I] 13.002 MEM/ 1000 13.003 ACC 3.000kH 10P% 13.004 MEME 13.005 MEMC kH 0.025U #! Test Tol 0.025, Sys Tol 3e-005, TUR 833.333 (>= 4.00). 14.001 HEAD { Load Regulation Test } 14.002 JMP 15.002 14.003 EVAL 15.001 DISP Remove the Clarostat from the AC terminal. 15.002 IEEE [@1281]RDG?[D2000][I] 15.003 MEME 15.004 JMPT 15.007 Validyne MC1-20:Twenty Channel Module Cas 15.005 DISP Set the Clarostat to 100 ohms and connect from one 15.005 DISP AC terminal to GRD. Make this connection on the LOAD 15.005 DISP board. 15.006 JMP 15.008 15.007 DISP Set the Clarostat to 1 ohm and connect from one 15.007 DISP AC terminal to GRD. Make this connection on the LOAD 15.007 DISP board. 15.008 IEEE [@1281]RDG?[D2000][I] 15.009 MEMC V 0.100% #! T.U.R. not calculated because System Uncertainty not available. 16.001 HEAD { Line Regulation Test } 16.002 JMP 17.001 16.003 EVAL 17.001 DISP Make sure the variac is set at 115 volts. 17.002 IEEE [@1281]RDG?[D2000][I] 17.003 MEME >1 17.004 DISP Set the variac to 104 volts. 17.005 IEEE [@1281]RDG?[D2000][I] 17.006 MEME 17.007 MEME <1 17.008 MEMC V 0.0100% 104VAC #! T.U.R. not calculated because System Uncertainty not available. 18.001 DISP Set the variac to 127 volts. 18.002 IEEE [@1281]RDG?[D2000][I] 18.003 MEME 18.004 MEME <1 18.005 MEMC V 0.0100% 127VAC #! T.U.R. not calculated because System Uncertainty not available. 19.001 DISP Set the variac to 115 volts. 19.002 HEAD { Frequency Load Regulation Test } 19.003 JMP 20.001 19.004 EVAL 20.001 IEEE [@1281]FREQ?[D2000][I] 20.002 MEM/ 1000 20.003 MEME #20.004 DISP Set the Clarostat to 100 ohms and connect from one #20.004 DISP AC terminal to GRD. 20.004 IEEE [@1281]FREQ?[D2000][I] 20.005 MEM/ 1000 20.006 MEMC kH 0.100% #! T.U.R. not calculated because System Uncertainty not available. 21.001 HEAD { Frequency Line Regulation Test } 21.002 JMP 22.001 21.003 EVAL 22.001 DISP Make sure the variac is set at 115 volts. 22.002 IEEE [@1281]FREQ?[D2000][I] 22.003 MEM/ 1000 22.004 MEME >1 22.005 DISP Set the variac to 104 volts. 22.006 IEEE [@1281]FREQ?[D2000][I] 22.007 MEM/ 1000 22.008 MEME 22.009 MEME <1 22.010 MEMC kH 0.0100% 104VAC #! T.U.R. not calculated because System Uncertainty not available. 23.001 DISP Set the variac to 127 volts. 23.002 IEEE [@1281]FREQ?[D2000][I] 23.003 MEM/ 1000 23.004 MEME 23.005 MEME <1 23.006 MEMC kH 0.0100% 127VAC #! T.U.R. not calculated because System Uncertainty not available. 24.001 DISP Set the variac to 115 volts. 24.002 DISP Turn off the variac. 24.003 DISP Disconnect all test equipment from the UUT. 24.004 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 12 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.